DMA 3200 Dynamic Mechanical Analyzer: High Force DMA

The DMA 3200 dynamic mechanical analyzer combines DMA and fatigue testing in one instrument. Moreover, it delivers up to 500 N of force. It is part of the TA Instruments ElectroForce range. Bargal represents TA Instruments and the ElectroForce line in Israel.

  • Force up to 500 N
  • DMA frequencies from 0.01 to 100 Hz
  • Forced convection oven from -150 to 600 °C
  • Fatigue tests up to 300 Hz
תיאור

The DMA 3200 dynamic mechanical analyzer is a high force instrument. It belongs to the ElectroForce range of TA Instruments. It combines DMA with fatigue and quasi-static testing. As a result, one platform handles small bars and larger samples. Moreover, it applies up to 500 N of force. In Israel, Bargal supplies the DMA 3200.

Who is the DMA 3200 for?

The DMA 3200 suits labs that must test stiff or large samples. It also suits teams that want DMA and fatigue data from one instrument. Typical users include:

  • Polymer and composite labs that test stiff or thick samples
  • Teams that also test larger samples or actual components
  • Labs that must understand when components fail
  • R&D groups that need DMA and tensile data on one platform

Applications

  • High force DMA: complex modulus, storage and loss moduli, and tan δ
  • Tension tests on thick samples, up to 8 mm in the LSO clamp
  • Fatigue testing: how and when materials, components or devices fail
  • Quasi-static tests, as well as ramp-to-failure tests
  • Temperature, frequency and strain sweeps, as well as time-temperature superposition

Why choose the DMA 3200 dynamic mechanical analyzer?

  • Frictionless ElectroForce linear motor, with up to 500 N of force
  • High-resolution optical displacement sensor, down to 1 nm
  • Interchangeable force sensors, so soft samples can use the optional 22 N sensor
  • Rigid test frame and air bearings, so very stiff samples give accurate results
  • Two ovens: a forced convection oven to 600 °C, and a Large Sample Oven
  • Air Chiller Systems, so you can cool below ambient without liquid nitrogen
  • WinTest control with TuneIQ, as well as TRIOS data analysis

Technical specifications

Parameter DMA 3200
Maximum force (DMA) 500 N
Maximum force (fatigue and quasi-static) 450 N
Minimum force (DMA) 0.2 N (0.025 N with the optional 22 N sensor)
Dynamic displacement range (DMA) ±0.0005 to ±6.5 mm
Displacement resolution 1 nm
Maximum acceleration 80 G
Frequency range (DMA) 0.01 to 100 Hz
Frequency range (fatigue and quasi-static) 0.00001 to 300 Hz
Modulus range (DMA) 10³ to 3 × 10¹² Pa
Modulus precision (DMA) ±1%
Tan δ sensitivity (DMA) 0.0001
Forced Convection Oven (FCO) temperature -150 to 600 °C (testing above 500 °C requires high temperature clamps)
FCO heating and cooling rate 0.1 to 60 °C/min
FCO isothermal stability ±0.1 °C
Large Sample Oven (LSO) temperature -150 to 315 °C
LSO heating and cooling rate 0.5 to 10 °C/min
LSO isothermal stability ±2 °C
Clamps Tension; compression; three-point bend; dual and single cantilever; shear sandwich; sample size limits differ by oven
Software WinTest control with TRIOS data analysis

 

Clamps, ovens and cooling

The DMA 3200 offers two ovens and many clamp types. Therefore, you can match the setup to your samples. In addition, all forced convection oven fixtures fit the Large Sample Oven with adapters.

Also, you can read the full details on the official TA Instruments DMA 3200 page.

 

Frequently asked questions

What is a DMA?

A DMA is a dynamic mechanical analyzer. It applies an oscillating stress or strain to a sample and measures the response. As a result, it gives modulus and damping (tan δ). Moreover, it works over a range of temperature, frequency or time.

What can the DMA 3200 dynamic mechanical analyzer measure?

It measures complex modulus, storage and loss moduli, and tan δ. It also runs fatigue and quasi-static tests.

Why does the DMA 3200 use high force?

High force allows larger samples and actual components. As a result, you can test thick samples in tension.

Can I cool below room temperature without liquid nitrogen?

Yes. Air Chiller Systems provide gas flow cooling without liquid nitrogen. In addition, the forced convection oven can reach -100 °C this way.

Can the DMA 3200 run fatigue tests?

Yes. It runs fatigue and ramp-to-failure tests. Moreover, you can use sine, triangle, square or ramp waveforms.

Which clamps are available?

Tension, compression, three-point bend, dual and single cantilever, and shear sandwich clamps. However, sample size limits differ by oven.

Is the DMA 3200 an ElectroForce product?

Yes. The DMA 3200 is part of the ElectroForce mechanical test instruments from TA Instruments. Moreover, it uses the patented ElectroForce linear motor. However, it is a different product from the ElectroForce 3200 Series III load frame.

 

Also see the Discovery DMA 850 dynamic mechanical analyzer.

For dimensional changes, see the Discovery TMA 450 thermomechanical analyzer.

 

Interested in the DMA 3200 dynamic mechanical analyzer? Send us an enquiry, and our team will respond quickly. We are also happy to discuss your samples and your application.

 

 

מנתח מכני דינמי DMA 3200 מקו ElectroForce של TA Instruments, עם כוח גבוה של עד 500 N ב-DMA ו-450 N בבדיקות עייפות, לדגימות גדולות וקשיחות – באספקת ברגל ציוד אנליטי בישראל.

צרו קשר

היי! שמחים שהגעתם לאתר שלנו! באתר יש כרגע מאות מוצרים ואנחנו מוסיפים מוצרים כל הזמן. אם לא מצאתם את מה שחיפשתם או שאין לכם כוח לחפש (קורה לכולנו), כתבו לנו כאן מה המוצר שאתם מחפשים ונשלח אליכם הצעת מחיר/נחזור אליכם

    דילוג לתוכן